In the framework of the ORAMED project (Optimization of RAdiation protection for MEDical staff), funded by the European Union Seventh Framework Programme, different studies were aimed at improving the quality of radiation protection in interventional radiology and nuclear medicine. The main results of the project were presented during a final workshop held in Barcelona in January 2011, the proceedings of which are available in the open literature. One of the ORAMED tasks was focused on the problem of eye-lens photon exposure of the medical staff, a topic that gained more importance especially after the ICRP decision to lower the limiting equivalent dose to 20 mSv per year. The present technical note has the scope, besides briefly summarising the physical reasons of the proposal and the practical implications, to provide, in tabular form, a set of air kerma to Hp(3) conversion coefficients based on the adoption of a theoretical cylindrical model that is well suited for reproduction of the mass and the shape of a human head.

Air kerma to HP(3) conversion coefficients for photons from 10 keV to 10 MeV, calculated in a cylindrical phantom.

Ferrari, P.;Mariotti, F.;Fantuzzi, E.;
2013-01-01

Abstract

In the framework of the ORAMED project (Optimization of RAdiation protection for MEDical staff), funded by the European Union Seventh Framework Programme, different studies were aimed at improving the quality of radiation protection in interventional radiology and nuclear medicine. The main results of the project were presented during a final workshop held in Barcelona in January 2011, the proceedings of which are available in the open literature. One of the ORAMED tasks was focused on the problem of eye-lens photon exposure of the medical staff, a topic that gained more importance especially after the ICRP decision to lower the limiting equivalent dose to 20 mSv per year. The present technical note has the scope, besides briefly summarising the physical reasons of the proposal and the practical implications, to provide, in tabular form, a set of air kerma to Hp(3) conversion coefficients based on the adoption of a theoretical cylindrical model that is well suited for reproduction of the mass and the shape of a human head.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12079/2939
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
social impact