A comparison was conducted, between 11 European National Metrology Institutes and EC-JRC, on measurement of 137Cs activity concentration in fume dust. As test material an activity standard produced from real contaminated fume dust was used. The standard material consisted of 13 cylindrical samples of compressed fume dust. The material contained 137Cs and 60Co of reference activity concentrations of (9.72±0.10)Bq/g and (0.450±0.018)Bq/g, respectively, for the reference date of 1 June 2013, determined using the comparison results. The organization and results of the intercomparison, as well as the process of obtaining reliable reference values are presented. © 2015 The Authors.

Interlaboratory comparison on 137Cs activity concentration in fume dust

Fazio, A.
2015-01-01

Abstract

A comparison was conducted, between 11 European National Metrology Institutes and EC-JRC, on measurement of 137Cs activity concentration in fume dust. As test material an activity standard produced from real contaminated fume dust was used. The standard material consisted of 13 cylindrical samples of compressed fume dust. The material contained 137Cs and 60Co of reference activity concentrations of (9.72±0.10)Bq/g and (0.450±0.018)Bq/g, respectively, for the reference date of 1 June 2013, determined using the comparison results. The organization and results of the intercomparison, as well as the process of obtaining reliable reference values are presented. © 2015 The Authors.
2015
Iodizing radiation measurements;Interlaboratory comparisons;Fume dust;Metrology
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12079/401
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
social impact