Nuclear radiation spectroscopy is now often relying on digital acquisition techniques. The present paper addresses the problem of analyzing the requirements of the digitizer in terms of sampling frequency and number of bits to minimize the energy resolution degradation caused by the digitizing process. The analysis is performed using synthetic pulses (with different amplitude and Full Width Half Maximum) of typical nuclear spectroscopy detectors and the pulse area as energy estimate. Additional relevant issues, such as the hardware architecture and the data throughput speed, are also discussed. © 2015 Published by Elsevier B.V.
Digital nuclear radiation spectroscopy: Hardware requirements to minimize energy resolution degradation
Belli, F.;Marocco, D.;Esposito, B.;Riva, M.
2015-01-01
Abstract
Nuclear radiation spectroscopy is now often relying on digital acquisition techniques. The present paper addresses the problem of analyzing the requirements of the digitizer in terms of sampling frequency and number of bits to minimize the energy resolution degradation caused by the digitizing process. The analysis is performed using synthetic pulses (with different amplitude and Full Width Half Maximum) of typical nuclear spectroscopy detectors and the pulse area as energy estimate. Additional relevant issues, such as the hardware architecture and the data throughput speed, are also discussed. © 2015 Published by Elsevier B.V.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.