The feasibility of a subminiature scintillation detector to be inserted in a catheter for lesion localization in nuclear medicine SPECT has been studied. Measurements on a simple laboratory setup have been performed and compared with Monte Carlo results. Further simulations, at 30keV and 140keV, concerning a configuration reproducing severe clinical conditions have shown poor lesion detectability. Several factors affecting the response have to be investigated to improve the capability of lesion localization characterizing such detector.

A subminiature scintillation detector for catheter operation

Burgio N.;Iurlaro G.;Santagata A.;Alonge G.
2006-01-01

Abstract

The feasibility of a subminiature scintillation detector to be inserted in a catheter for lesion localization in nuclear medicine SPECT has been studied. Measurements on a simple laboratory setup have been performed and compared with Monte Carlo results. Further simulations, at 30keV and 140keV, concerning a configuration reproducing severe clinical conditions have shown poor lesion detectability. Several factors affecting the response have to be investigated to improve the capability of lesion localization characterizing such detector.
2006
978-981-256-798-7
978-981-277-367-8