In the framework of the Italian Space Agency (ASI) Supported Irradiation Facilities (ASIF) program, a campaign of dosimetric measurements was conducted to perform an intercalibration of the Calliope gamma irradiation facility and the REX electron beam. Initially, up-to-date calibration curves for alanine-EPR dosimeters were obtained at the Calliope facility using the absolute Fricke dosimeter. Subsequently, alanine dosimeters were irradiated at multiple positions in the REX chamber by varying distances from the electron source. The resulting dose rate distribution within the chamber was determined. These measurements are essential for optimizing REX irradiation conditions to be used in different applications.

Nel contesto del programma dell’Agenzia Spaziale Italiana (ASI) ASI Supported Irradiation Facilities (ASIF), è stata condotta una campagna di misure dosimetriche per eseguire l'intercalibrazione tra l’impianto di irraggiamento gamma Calliope e il fascio di elettroni di REX. Inizialmente, sono state ottenute le curve di calibrazione aggiornate per i dosimetri alanina-EPR presso la facility Calliope utilizzando il dosimetro assoluto Fricke. Successivamente, i dosimetri ad alanina sono stati irraggiati in più punti all'interno del bunker di irraggiamento di REX, variando le distanze dalla sorgente di elettroni. È stata così determinata la distribuzione dell’intensità di dose all'interno della camera. Tali misure sono essenziali per ottimizzare le condizioni di irraggiamento all’interno del bunker di REX per applicazioni in diversi campi.

Dosimetric measurements for precise dose rate distribution mapping: ENEA Calliope gamma radiation facility and REX electron beam intercalibration

Di Sarcina, I.;Verna, A.;Cemmi, A.;Ampollini, A.;Astorino, M. D.;Bazzano, G.;Ronsivalle, C.;Trinca, E.;
2025-01-01

Abstract

In the framework of the Italian Space Agency (ASI) Supported Irradiation Facilities (ASIF) program, a campaign of dosimetric measurements was conducted to perform an intercalibration of the Calliope gamma irradiation facility and the REX electron beam. Initially, up-to-date calibration curves for alanine-EPR dosimeters were obtained at the Calliope facility using the absolute Fricke dosimeter. Subsequently, alanine dosimeters were irradiated at multiple positions in the REX chamber by varying distances from the electron source. The resulting dose rate distribution within the chamber was determined. These measurements are essential for optimizing REX irradiation conditions to be used in different applications.
2025
Nel contesto del programma dell’Agenzia Spaziale Italiana (ASI) ASI Supported Irradiation Facilities (ASIF), è stata condotta una campagna di misure dosimetriche per eseguire l'intercalibrazione tra l’impianto di irraggiamento gamma Calliope e il fascio di elettroni di REX. Inizialmente, sono state ottenute le curve di calibrazione aggiornate per i dosimetri alanina-EPR presso la facility Calliope utilizzando il dosimetro assoluto Fricke. Successivamente, i dosimetri ad alanina sono stati irraggiati in più punti all'interno del bunker di irraggiamento di REX, variando le distanze dalla sorgente di elettroni. È stata così determinata la distribuzione dell’intensità di dose all'interno della camera. Tali misure sono essenziali per ottimizzare le condizioni di irraggiamento all’interno del bunker di REX per applicazioni in diversi campi.
Gamma irradiation
Electron irradiation
Alanine-EPR dosimetry
Fricke dosimetry
Irraggiamento gamma
Irraggiamento con elettroni
Dosimetria ad alanina
Dosimetria Fricke
File in questo prodotto:
File Dimensione Formato  
RT-2025-03-ENEA.pdf

accesso aperto

Descrizione: Documento completo
Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 2.22 MB
Formato Adobe PDF
2.22 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12079/80107
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
social impact