This report presents a study on the characterization of thermoluminescence dosimeters (TLDs) using 60Co gamma radiation at the Calliope Facility and the National Institute of Ionizing Radiation Metrology (INMRI) of the ENEA Casaccia Research Center. It covers thermoluminescence theory and the TLD reading process, followed by dose evaluation and calibration techniques. The dosimeters used, particularly TLD 100, are described in detail, with an explanation of the irradiation process and calibration procedure. Results from low-dose calibration and comparisons with previous INMRI calibrations are presented. Additionally, comparisons between Fricke and TLD dosimetry for short irradiation times and low dose rates are included, along with discussions on shielding. This study aims to enhance the accuracy of dose measurements, especially for small organic samples for possibile radiobiological applications.

Questo report presenta uno studio sulla caratterizzazione dei dosimetri a termoluminescenza (TLD) utilizzando radiazioni gamma da Cobalto-60 presso la facility Calliope e l'Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (INMRI) del Centro di Ricerca ENEA Casaccia. Vengono trattati la teoria della termoluminescenza, il processo di lettura del TLD e le tecniche di valutazione della dose e le relative tecniche di calibrazione. I dosimetri utilizzati, in particolare i TLD 100, la procedura di irraggiamento e calibrazione sono descritti nel dettaglio. Vengono presentati i risultati della calibrazione a bassa dose e un confronto con le calibrazioni precedenti effettuate presso l'INMRI. Inoltre, vengono forniti confronti tra dosimetria Fricke e TLD per brevi tempi di irraggiamento e basse dosi, e vengono introdotti i risultati ottenuti nel caso di schermatura. Questo studio mira a migliorare l'accuratezza delle misure della dose, soprattutto per campioni organici di dimensioni ridotte per possibili applicazioni in radiobiologia.

Thermoluminescence dosimeters characterization with 60Co gamma radiation at ENEA Casaccia Research Center

Di Sarcina, I.;Verna, A.;Cemmi, A.;Mansi, E.;
2025-01-01

Abstract

This report presents a study on the characterization of thermoluminescence dosimeters (TLDs) using 60Co gamma radiation at the Calliope Facility and the National Institute of Ionizing Radiation Metrology (INMRI) of the ENEA Casaccia Research Center. It covers thermoluminescence theory and the TLD reading process, followed by dose evaluation and calibration techniques. The dosimeters used, particularly TLD 100, are described in detail, with an explanation of the irradiation process and calibration procedure. Results from low-dose calibration and comparisons with previous INMRI calibrations are presented. Additionally, comparisons between Fricke and TLD dosimetry for short irradiation times and low dose rates are included, along with discussions on shielding. This study aims to enhance the accuracy of dose measurements, especially for small organic samples for possibile radiobiological applications.
2025
Questo report presenta uno studio sulla caratterizzazione dei dosimetri a termoluminescenza (TLD) utilizzando radiazioni gamma da Cobalto-60 presso la facility Calliope e l'Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (INMRI) del Centro di Ricerca ENEA Casaccia. Vengono trattati la teoria della termoluminescenza, il processo di lettura del TLD e le tecniche di valutazione della dose e le relative tecniche di calibrazione. I dosimetri utilizzati, in particolare i TLD 100, la procedura di irraggiamento e calibrazione sono descritti nel dettaglio. Vengono presentati i risultati della calibrazione a bassa dose e un confronto con le calibrazioni precedenti effettuate presso l'INMRI. Inoltre, vengono forniti confronti tra dosimetria Fricke e TLD per brevi tempi di irraggiamento e basse dosi, e vengono introdotti i risultati ottenuti nel caso di schermatura. Questo studio mira a migliorare l'accuratezza delle misure della dose, soprattutto per campioni organici di dimensioni ridotte per possibili applicazioni in radiobiologia.
Gamma irradiation
60Co
Thermoluminescence dosimetry
Calibration
Irradiamento gamma
Cobalto 60
Dosimetria
Calibrazione
File in questo prodotto:
File Dimensione Formato  
RT-2025-04-ENEA.pdf

accesso aperto

Descrizione: Documento completo
Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 2.4 MB
Formato Adobe PDF
2.4 MB Adobe PDF Visualizza/Apri