In 2014, comparisons of activity measurements of 99mTc using the Transfer Instrument of the International Reference System (SIRTI) took place at the VNIIM (Russian Federation) and at the ENEA-INMRI (Italy), with comparison identifier BIPM.RI(II)-K4.Tc-99m. Ampoules containing about 32 kBq of 99mTc solutions were measured in the SIRTI for three half-lives. There are now nine results in the BIPM.RI(II)-K4.Tc-99m comparison. This comparison is linked to the BIPM.RI(II)-K1.Tc-99m comparison and the key comparison reference value (KCRV) of the BIPM.RI(II)-K1.Tc-99m has been updated to include eligible results from BIPM.RI(II)-K4.Tc-99m. The degrees of equivalence between each equivalent activity measured in the SIR or the SIRTI and the KCRV have been calculated and the results are given in the form of a table. A graphical presentation is also given.
Titolo: | Activity measurements of the radionuclide 99mTc for the VNIIM, Russian Federation and ENEA-INMRI, Italy, in the ongoing comparison BIPM.RI(II)-K4.Tc-99m and KCRV update in the BIPM.RI(II)-K1.Tc-99m comparison |
Autori: | |
Data di pubblicazione: | 2016 |
Rivista: | |
Handle: | http://hdl.handle.net/20.500.12079/1748 |
Appare nelle tipologie: | 1.1 Articolo in rivista |