A calibration procedure is proposed for ITER Radial Neutron Camera that relies on embedded sources, reference ITER pulses and cross-calibration with ITER fission chambers and activation system coupled to Monte Carlo simulations of radiation transport. The proposed procedure would allow to measure the neutron emissivity profile and of the fusion power with 10 % accuracy and precision, a time resolution of 10 ms and a spatial resolution of a/10 for ITER entire life-time.

Strategy and guidelines for the calibration of the ITER Radial Neutron Camera

Marocco D.;Moro F.;Esposito B.;Podda S.;
2019-01-01

Abstract

A calibration procedure is proposed for ITER Radial Neutron Camera that relies on embedded sources, reference ITER pulses and cross-calibration with ITER fission chambers and activation system coupled to Monte Carlo simulations of radiation transport. The proposed procedure would allow to measure the neutron emissivity profile and of the fusion power with 10 % accuracy and precision, a time resolution of 10 ms and a spatial resolution of a/10 for ITER entire life-time.
2019
Calibration; ITER; Radial Neutron Camera
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12079/51951
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
social impact