A key comparison has been made between the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty for the comparison of 1.9 parts in 103. No significant trend with radiation quality is observed. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.

Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in low-energy x-rays

Toni Maria Pia;Pinto Massimo;Bovi Maurizio;Cappadozzi Gianluca;Silvestri Claudia
2011-01-01

Abstract

A key comparison has been made between the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty for the comparison of 1.9 parts in 103. No significant trend with radiation quality is observed. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.
2011
dosimetry
INMRI
key comparison
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12079/55563
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
social impact