In a medical accelerator, real-time monitoring systems of the beam and dose delivered to the patient are mandatory. In this work, we present a compact current profile detector that has been designed and tested in the framework of the TOP-IMPLART (Intensity Modulated Proton Linear Accelerator for RadioTerapy) project. This project foresees the realization of a proton linear accelerator, currently under construction at ENEA Frascati, for proton therapy applications. The linac produces a pulsed proton beam with 3 μs duration at 50 Hz repetition rate with a pulse current between 0.5 and 50 μA. A large dynamic range and spatial constraints make the use of usual noninterceptive beam diagnostics unfeasible. Therefore, the use of a beam current monitor based on a passive RF cavity working in the TM010 mode has been proposed. This paper reports the electromagnetic design of the device guided by a simplified analytical model. A prototype of such a device has been realized, characterized, and tested on the linac with a 35 MeV beam varying the beam current. The test results in air and in vacuum, together with the signal detection systems used, are presented.

Design and test of a compact beam current monitor based on a passive RF cavity for a proton therapy linear accelerator

Ampollini A.;Bazzano G.;Nenzi P.;Ronsivalle C.;Picardi L.
2021-01-01

Abstract

In a medical accelerator, real-time monitoring systems of the beam and dose delivered to the patient are mandatory. In this work, we present a compact current profile detector that has been designed and tested in the framework of the TOP-IMPLART (Intensity Modulated Proton Linear Accelerator for RadioTerapy) project. This project foresees the realization of a proton linear accelerator, currently under construction at ENEA Frascati, for proton therapy applications. The linac produces a pulsed proton beam with 3 μs duration at 50 Hz repetition rate with a pulse current between 0.5 and 50 μA. A large dynamic range and spatial constraints make the use of usual noninterceptive beam diagnostics unfeasible. Therefore, the use of a beam current monitor based on a passive RF cavity working in the TM010 mode has been proposed. This paper reports the electromagnetic design of the device guided by a simplified analytical model. A prototype of such a device has been realized, characterized, and tested on the linac with a 35 MeV beam varying the beam current. The test results in air and in vacuum, together with the signal detection systems used, are presented.
2021
Humans
Particle Accelerators
Protons
Proton Therapy
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12079/60924
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
social impact