A comparison of the activity concentration of an 55Fe solution was organized by the POLATOM under the auspices of the CCRI(II). The solution was sent to thirteen laboratories; ready-to-measure sources were also prepared using the same 55Fe solution in diffusive (polyethylene) vials and sent to twelve laboratories. Twelve measurement reports were received. The results, as well as relevant information on the measurement techniques, are presented and analysed in this paper. The differences in the results obtained from measurements in non-diffusive (glass) and diffusive vials are indicated.

Final report of the key comparison CCRI(II)-K2.Fe-55.2019 of activity concentration measurements of a 55Fe solution

Broda R.;Capogni M.;Carconi P.;De Felice P.;Fazio A.;
2021-01-01

Abstract

A comparison of the activity concentration of an 55Fe solution was organized by the POLATOM under the auspices of the CCRI(II). The solution was sent to thirteen laboratories; ready-to-measure sources were also prepared using the same 55Fe solution in diffusive (polyethylene) vials and sent to twelve laboratories. Twelve measurement reports were received. The results, as well as relevant information on the measurement techniques, are presented and analysed in this paper. The differences in the results obtained from measurements in non-diffusive (glass) and diffusive vials are indicated.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12079/66149
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
social impact