This work presents new calculations of conversion coefficients (CCs) from total air kerma to dose-equivalent quantities H’(0.07), H’(3), H*(10), Hp(0.07), Hp(3), and Hp(10) for area and personal dosimetry for mono-energetic photons in the energy range from 2 keV to 50 MeV, assuming secondary charged particle equilibrium. Calculations using the Monte Carlo N-Particle® (MCNP) code were performed for a large number of photon energies with the aim of preventing errors resulting from possible improper interpolation between currently available sparsely spaced CC values, when an average CC value over a photon fluence spectrum needs to be determined. The CC values were compared with the values published in the ISO 4037-3:2019 standard. A very close agreement was achieved for the majority of CCs. Larger discrepancies were found for some CCs, often for low photon energies or large angles of radiation incidence, which were taken from older publications or when CC values were interpolated or extrapolated. Furthermore, some differences were found in the MeV energy range, which are significant for dosimeter calibrations, e.g. the presented values of CC to H*(10) for the main photon energies of 137Cs and 60Co radionuclides are both lower by 2.8%. Finally, it was found that the values of CCs to Hp(0.07; E, α)slab given in ISO 4037-3:2019 were not taken correctly from the source publication. In conclusion, the CC values given in ISO 4037-3:2019 should be updated in view of the results obtained.

Re-calculation of air kerma to dose-equivalent conversion coefficients for mono-energetic photons

Pinto M.;
2025-01-01

Abstract

This work presents new calculations of conversion coefficients (CCs) from total air kerma to dose-equivalent quantities H’(0.07), H’(3), H*(10), Hp(0.07), Hp(3), and Hp(10) for area and personal dosimetry for mono-energetic photons in the energy range from 2 keV to 50 MeV, assuming secondary charged particle equilibrium. Calculations using the Monte Carlo N-Particle® (MCNP) code were performed for a large number of photon energies with the aim of preventing errors resulting from possible improper interpolation between currently available sparsely spaced CC values, when an average CC value over a photon fluence spectrum needs to be determined. The CC values were compared with the values published in the ISO 4037-3:2019 standard. A very close agreement was achieved for the majority of CCs. Larger discrepancies were found for some CCs, often for low photon energies or large angles of radiation incidence, which were taken from older publications or when CC values were interpolated or extrapolated. Furthermore, some differences were found in the MeV energy range, which are significant for dosimeter calibrations, e.g. the presented values of CC to H*(10) for the main photon energies of 137Cs and 60Co radionuclides are both lower by 2.8%. Finally, it was found that the values of CCs to Hp(0.07; E, α)slab given in ISO 4037-3:2019 were not taken correctly from the source publication. In conclusion, the CC values given in ISO 4037-3:2019 should be updated in view of the results obtained.
2025
area dosimetry
conversion coefficients
ISO 4037 standard
MCNP
Monte Carlo
personal dosimetry
total air kerma
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12079/89247
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
social impact